TECHNICAL DATA & INSTALLATION
General Information
1.1 General Description
The instructions purpose is to describe performance, installation, debugging, operation, maintenance and troubleshooting of PAX-F Series Shadowless Operation Lamp. There is written description on wall and mobile products. WARNING: The device is only used for operation lighting in the medical units, and is forbidden for other purposes. Any change of the device or parts will be forbidden without the company's written authorization.
1.2 Definition
The following are the meanings of identifications used in the instructions: NOTICE: Indicating that there are important contents which should be attached importance to. CAUTION: Appearing before some operations, and once these provisions are disobeyed, the device may be damaged. WARNING: Appearing before some operations, and once these provisions are disobeyed, the device may be damaged and personal safety may be endangered.
2. Product Introduction
2.1 Product Characteristics
- Configured professional balancer will make lamp body's balance stable and reliable.
- High-emission reflection device made up of imported materials may ensure taht illumination degree of sahdowless operation lamp is permanently efficient.
- Streamline design with lower air damp coefficient may be used in high-demand purifying operating room.
- Suitable light and color temperature may provide high-quality optics performance.
- Shadowless operation lamp made up of lightweight materials may provide more portable and more flexible operation.
- Lightness-adjusting handle taht may be disinfected will guarantee facula micro-adjustment performed in the process of the operation to keep in asepsis status.
- Auto-monitoring computer program will track the whole process of using the device and switch damged bulbs in time.
- Soft starting system extends bulb's service life.
- Independent electrical system of combination products will improve device's reliability.
| TECHNICAL DATA |
CONTENTS |
F700/700, F700, F700/500, F500/500, F500 |
Power Voltage |
220V, 10AC |
Input Power |
50Hz |
Bulb Power |
600 VA, 300VA, |
Main Fuse |
150W/150W, 150W |
MECHANICAL CHARACTERISTICS |
CONTENTS |
ALL MODELS |
Revolver rotating around central axis |
360 |
Balancer rotating around the revolver |
320 |
Balancer running up & down |
80 |
Lamp head rotating around the balancer |
320 |
Lamp head rotatig among connecting tubes |
180 |
Lamp body rotating around the connecting tubes |
180 |
PERFORMANCE CHARACTERISTICS |
CONTENTS |
EACH LIGHT SOURCE |
Rated Light |
F700, 50000-140000lx, F500, 40000-100,000lx |
Color Temperature |
4300 |
Comforty Description
3.1 Device Type
Anti-electric shock type |
I |
Anti-electric shock degree |
B |
Running mode |
Continual running |
Protection degree |
Universal device not preventing liquid from entering |
Using location |
Can't be used in flammable and exploxive environment |
3.2 Product Standard
Performing the standards of eneterprises registered product Q/TCGU 5, PAX-F/K Series of Shadowless Operation
INSTALLATION
Specification before installing shadowless operation lamp
1. BASIC REQUIREMENT
Shadowless operation is a special device hung in the roof of operating room, and the security of basic installation is specially important. The product's basic requirement is to be capable of loading 500 kg static load, have enough rigidity, and insure bearable load and rigidity. Standard height of operating room is 3 meters
Requirement description |
Design |
2.1 Dimensions of preset plywood Four 14 holes are processed on the plywood, and the hole positions are evenly distributed among 240 mm diameters. Plywood thickness 12 mm, and the material is No.45 Steel. Plywood external diameter 360 mm, and may also be processed into square dimension 360 mm.
|
|
2.2 Ceiling plywood-type
A pice of plywood is placed respetively at the top and bottom of the ceiling and it should be fixed with M12 bolt, nut, spring gasket and flat gasket. M12 bolt, with screw thread should be longer 60 mm than the plywood.
|
|
2.3 Ceiling pre-embedding
When construction is being performed on the ceiling, four ring bolts (M12) should be embedded into the ceiling, and ring bolts must be jointed with the steel bar. Position and dimension of ring bolts should be decided according to the requirements in Figure 1. M12 bolt (with screw thread) should be longer 60 mm than the plywood.
|
|
2.4 Beam joining mode
The plywood should be jointed with the angle iron, and the holes should be drilled in the beam and then M16 bolts and nuts should be fixed in the beam. Four M12 60 bolts should be jointed on the plywood, and their dimensions should be processed according to the requirements in Table 2.1
|
|
2.5 Installation modes of channel steel. Two pieces of channel steel will be installed on the wall, and plywood will be fixed with M12 bolt. M12 bolt (with screw thread) should be longer 60 mm than the plywood. Notice: The installation of channel steel must ensure taht shadowless operation lamp won't shake when it is used.
|
|
2.6 Basic currency
Soleplate's dimension of the company is identical with other domestic companies. If the uer replaces the product, foundation board needn't be altered. For safety, the user should carefully check the instensity and status of original foundation board. If the dimension of existing foundation board doens't conform to the dimension required by the company, the company may customize communicating board for the users
|
2.7 Solutions of super-high operating room
If the height of operating room is more than 3.2 meter, height-extending shelf must be installed on the plywood, and the users may order the shelf in the company.
|
2.8 Power cabling requirements
Power requirement for shadowless operation lamp is AC 220V, 10, 50Hz. The users may install power switch 10A and fuse 6.3A on the metope of operating room, and the power switch must cut off phase line and neutral line of the power simultaneously. Cabling position of pwer supply cords should be from power switch to foundation board. Secure capacity of power supply cords is 15A. WARNING: grounding cords of the power part must be reliable. Independent safety grounding cords must be connected, and be joined with protection grounding port of shadowless operation lamp.
|
INSTALLATION OF SHADOWLESS OPERATION LAMP
1 Packing box
PAX-F Series of shadowless operation lamp of the company has two specifications including F700/700 double-lamp and F700 single-lamp
Shipping list for products with above-mentioned specifications is follows:
Product model |
Packing box's name |
Amount |
F700/700 |
Double revolver, balance |
F700 lamp head box |
|
|
F700 |
Single revolver, balancer |
F700 lamp head box |
|
|
2. Opening the box
The user should check the packing box according to ordered product model. If the amount is missing, please make contact with the transportation dealer immediately. Open the revolver, balancer and lamp. Please check carefully the parts according to shipping list in the packing box. If there are missing parts, please make contact with the company immediately.
3. Device installation
Warning: Check foundation board's status once again, and make sure its security. Check electrical connection and safety fully, and make sure its security. The installation must be performed strictly according to the following requirements and procedures.
Installation Description |
Design |
3.1 Installing the soleplate
Screw four M12 into the foundation soleplate; embed four holes of installation soleplate into 4 pieces of screw plat gasket, spring gasket and nuts. Adjust bottom nuts so as to make installation soleplate keep horizontal. Scre tight the above 4 nuts.
|
 |
3.2 Installing the revolver
Screw four M12 nuts into the installation soleplate; embed four holes of revolver chassis into 4 pieces of screw of foundation soleplate, and then screw plat gasket, spring gasket and nuts. Adjust bottom nuts so as to make the revolver's chassis keep horizontal, and screw tight nuts.
|
 |
3.3 Installing the balancer 1
Connect electrical wires of bending pipes with electrical wires of balancer firmly via connection pole. Notice: Electrical wires must be connected with the colors aligned and the connection must be reliable.
|
 |
3.4 Installing Balancer 2
Dismantle 4 bolts of the balancer, push the balancer into the axis hole of the revoler, and screw 4 bolts into the screw.
|
 |
3.5 Installing the lamp body
Connect to the electrical wires of lamp bending pipes and balancer with the connection pole according to the color, cross connection pipes of lamp bending pipes into the balancer. fix the connection with 4 bolts. Notice: Electrical wires must be connected with the colors aligned and the connection must be reliable.
|
 |
3.6 Installing light source parts 1
Take out the bulbs of accessories bag, and insert them into the lamp holder. Examine whether electrical connection is reliable. Press down handle button of disinfection handle, and take down disinfection handle.
|
 |
3.7 Installing light source parts 2
Aligne orientation nail of light source part with the flute of lamp head plastic gasket, and push into the loght source part upwards.
|
 |
3.8 Installing light source parts 3
Screw tight three connection bolts of light source part with screw arbor.
|
 |
3.9 Installing light source parts 4
Align the flute of disinfection handle with handle button, and push into the handle axis of light source part. When hearing a cracking sound, disinfection handle has been oriented.
|
 |
3.10 Disassembly of balancer's orientation nail after the lamp body has been properly installed, open the plastic covers on both sides of the balancer, and dismantle one bolt and right-angle orientation nail at the bottom of the balancer. Fold the plastic covers. Warning: Disassembly of orientation nail must be performed after the lamp body is installed and without load in the balancer. Strong compression strength will make the balancer sring up rapidly. and therefore cause the damage of balancer parts.
|
 |
3.11 Installing the top cover
Connect electrical board of the revolver with network power firmly according to the wiring diagram, and connect protection grounding wires. Loose the bolts on top cover's fixing gasket, place the top cover on the revolver, and then join the top cover with 6 bolts. Fold the top cover's fixing gasket into the top cover's holes. Push up the top cover revolver's chassis or ceilling. Screw tight inner hexagon bolt of top cover fixing gasket
|
 |
4. Electrical Connections
See attached pages in the instructions for the details.
4. Debugging of shadowless operation lamp
Contents |
Design |
1. Horizontal asjustment
Rectify horizontal state of the revolver with the gradienter. If there is no gradienter, loosen the revolver's damper, and spread fully the lamp to the biggest dimension. Screw up the soleplates's nuts with the lamp in inclining direction until each running part of the lamp has basically been stooped.
|
|
2. Damp adjustment
After horizontal adjustment, screw tight the damper until appropriate degree. Make the lamp orient in any position, the lamp won't move automatically without outside force.
|
 |
3. Balancer adjustment
Open the plastic cover in the dront of the balancer, and pull the lamp head so as to make adjustment nuts of the balancer. Insert the adjustment pole into the adjustment nuts, and adjust the balancer's pull until appropriate degree. Close the plastic cover. Notice: The balancer's pull has been adjusted before leaving the factory, and therfore genarally the users don't need to make any adjustment.
|
 |
4. Lamp body rotating adjustment
Rotate rubber gasket of lamp body round pipe,a nd adjust inner hexagon bolt with inner hexagon wrench until appropriate rotating force. Notice: The lamp body's rotating pull has been adjusted before leaving the factory, and therefore generally the users don't need to make any adjustment.
|
 |
5. Rotating adjustment of lamp body connecting tube. Adjust inner hexagon wrench until appropriate rotating force. Notice: The lamp body's rotating pull has been adjusted before leaving the factory, and therefore generally the users dont need to make any adjustment.
|
 |
Troubleshooting of shadowless operation lamp
1. Troubleshooting performed by the users
Phenomenon |
Possible reason |
Solution |
| Power indicator light on the control panel isn't on |
Power isn't switced on |
Turn on power's switch on the wall |
| |
Power board's elecrical wires are cut off |
Reconnect power board's electrical wires |
| Power indicator light on the control panel isn't on |
Fuse is burned out |
Replace the fuse |
| Light source of the shadowless operation lamp is glittering |
Network power's voltage is too low |
Install balanced power supply |
| Bulb indicator light on the control panel is glittering |
Bulb is burned out |
Replace the bulb |
| Balancing force is too tight or too loose |
Balancer's spring is too big or too small |
Adjust balancer's spring |
| Revolver can't be oriented |
| Revolver's chassis isn't horizontal |
| Damper is loose |
|
| Adjust chassis's nuts |
| Screw tight the damper |
|
| Lamp body can't be oriented |
The bolts of lamp body rubber gasket are loose |
Screw tight the bolts |
Above-mentioned troubleshooting operation should be performed according to related chapters such as Chapter 3 Installation of Shadowless Operation Lamp Chapter 4 Debugging of Shadowless Operation Lamp. Warning: When performing the troubleshooting, network power must be cut off.
2. Replace the fuse
Open the revolver's top cover, and replace the fuse on power board. The operation should be performed accroding to related chapters and attached pages in Chapter 3 Installation of Shadowless Operation Lamp.
Notice: Fuse may be ordered in the company.
Warning: When replacing the fuse, network power must be cut off.
3. Replace the bulb
Take down disinfection handle, screw open three connecting bolts, pull out light source parts, and then replace the bulb. The operation should be performed according to related sections in Chapter 3 Installation of Shadowless Operation Lamp. Specification 700 lamp's bulbs; OSRAM 64642 24V 150W
Notice: Bulb may be ordered in the company.
Warning: Network power must be cut off, the bulb can't be replaced until it is cooling.
4. Information maintenance dealer or the company
If shadowless operation lamp can't function properly after performing the troubleshooting, the details should be informed timely to local maintenance dealer authorized by the company or maintenance department. And we will get rid of the faults within shortest time.
|